Soft X-ray Spectrometer Operation at the National Ignition Facility

by D. Lee, M. B. Schneider, R. Turner, F. Weber, K. Campbell, D. Pellinen, M. Rhodes, O. Landen, V. Tran, E. Dewald, J. Schein ยท 2006

ISBN:  Unavailable

Category: Unavailable

Page count: 16

Radiation drive diagnostics during the NIF early light campaign was supported by an 18 channel soft x-ray spectrometer (Dante). In order to achieve a measurement accuracy of 2% in radiation temperature absolute calibration of the individual channels was necessary and signal distortion through long transmission lines had to be compensated for as well. For fast signals the signal attenuation due to the long (50m) cables amounted to {approx} 20% {at} 100MHz, which was corrected by a cable compensation in the frequency domain. The varying effects of cable distortion for a variety of signals between 1ns and 9ns in length were evaluated and corrections were applied. Results of the thus calculated temperatures of the NEL campaign will be presented compared to LASNEX predictions, showing agreement in peak radiation temperature within less than 2%.